Technical data
STENT | |
---|---|
Stent Material | Cobalt chromium, L-605 |
Strut thickness | ø 2.25 - 3.0 mm: 60 μm (0.0024”); ø 3.50 - 4.0 mm: 80 μm (0.0031”) |
Passive coating | proBIO (Amorphous Silicon Carbide) |
Active coating | BIOlute bioabsorbable Poly-L-Lactide (PLLA) eluting a limus drug |
Drug dose | 1.4 μg/mm2 |
Delivery system | |
---|---|
Catheter type | Rapid exchange |
Recommended guide catheter | 5F (min. I.D. 0.056”) |
Guide wire diameter | 0.014” |
Usable catheter length | 140 cm |
Balloon material | Semi crystalline polymer material |
Coating (distal shaft) | Hydrophilic |
Coating (Proximal shaft) | Hydrophobic |
Marker bands | Two swaged platinum-iridium markers |
Lesion entry profile | 0.017” |
Distal shaft diameter | 2.7F: ø 2.25 - 3.0 mm; 2.9F: ø 3.5 - 4.0 mm |
Proximal shaft diameter | 2.0F |
Nominal pressure (NP) | 10 atm |
Rated burst pressure (RBP) | 16 atm |
Storage | |
Use Before Date (UBD) | 24 months |
Temperature | Between 15°C (59°F) and 25°C (77°F), short term excursions between 10°C (50°F) and 40°C (104°F) are allowed |